Bilevel Positive Airway Pressure (BPAP) System
Product details
- Bilevel Positive Airway Pressure (BPAP) System
Company announcement
React Health (3B Medical, Inc.) has issued a voluntary medical device correction for specific Luna G3 BPAP System (Model LG3800-30VT) devices.
The correction addresses a data handling issue that can lead to incorrect therapy pressure statistical data when the device is operated with Targeted VT enabled.
This issue affects devices distributed in the United States and Puerto Rico.
Fifteen complaints and two serious adverse reports have been received in connection with this problem. No deaths have been reported.
The company announced this on November 17, 2026, with the FDA publishing the alert on November 19, 2026.
Risk:
The hazardous condition requires the device to be operated with Targeted VT enabled and monitored remotely via either data downloaded to the SD card or via wireless communication to the cloud-based management system (React Health Connect). The consequence of the inaccurate pressure data in historical therapy reports could lead a clinician to adjust device settings based on inaccurate therapy pressure data, which could result in delayed therapy until the setting is corrected. Design features that reduce the risk of this issue include the device adjusting pressure delivered to the patient within the clinician defined minimum and maximum pressures delivering a tidal volume equal to the clinician set tidal volume, within tolerance.
Impacted states: Nationwide
Products:
- Luna® G3 BPAP System (Model LG3800-30VT)
Contact
For more information, call 863-226-6285 or email 3B-Recall@reacthealth.com.